Nano Spray Dryer B-90
![]() | World novelty for fine particles from minimal sample quantities at high yields! |
Open mode for aqueous samples and not oxygen sensitive materials
Drying with air for aqueous solutions with < 20 % organic solvents. The Dehumidifier B-296 is used for reproducible conditioning of the inlet drying air humidity.
Compressed air is used as a standard drying gas. The air maintenance unit is additionally recommended to assure oil free and particle free gas.
Closed mode for organic solvents and oxygen sensitive materials
Safe drying of pure organic solvents with automatic control of the O2 concentration in the circuit below 4 %. N2 and CO2 are used to control the efficiency of the electrical particle collector.
Drying of oxygen sensitive substances or flammable powders under N2 and CO2 atmosphere. The B-296 in a closed loop enables the safe use of water-organic solvent mixtures with up to 50 % solvent.