Pump Systems
![]() | Pump Systems - Powerful and chemical resistant |
Flexible and modular - Optimal combination for solvent delivery in Flash - and Preparative HPLC Chromatography
A robust and pulse free pump system is at the heart of Buchi Sepacore®. A pressure range up to either 10 bar or 50 bar gives optimum separation results for a broad range of applications. The pump modules can be controlled by three different units:
- Pump Controller C-610 for isocratic separation up to 10 bar
- Pump Manager C-615 for isocratic and gradient separation up to 50 bar
- Control Unit C-620 as the interface between the PC and all components in the Sepacore® system. The unit is included in the SepacoreControl package.
Pump Controller C-610 - Isocratic Flash Chromatography
Pump Controller C-610:
The Pump Controller C-610 for one Pump Module C-601 is designed for isocratic separations up to 10 bar with a flowrate of max. 250 ml/min. The flow rate can be easily adjusted by turning a knob and is indicated by a large illuminated LCD-display. Delivered with a overpressure sensor for maximum safety.
Pump System:
The Pump Controller C-610 with a Pump Module C-601 is used for fast isocratic Flash separations. No programming is needed. The system can be run by two buttons and one knob.
Pump Manager C-615 - Isocratic and gradient separations in Flash- and Preparative Chromatography
Pump Manager C-615
Designed for both isocratic and gradient separations up to 50 bar with a flowrate of max. 250 ml/min. Fast operation, easy programming and a large graphical display allows a quick and easy set up. Running time, solvent consumption and actual pressure are shown during a separation. The unit has Input/Outputs for 2 solvent valves and level sensors and includes a pressure sensor and mixing chamber.
Isocratic Pump System:
The Pump Manager C-615 with one Pump Module C-601/C-605 is used for isocratic Flash separations. This combination allows exploration into the features of the Pump Manager C-615 for solvent selection, timed runs and solvent level control.
Gradient Pump System:
The Pump Manager C-615 with two Pump Modules C-601/C-605 performs binary solvent gradients. The efficient solvent mixing under pressure and the pulsation free solvent flow eliminate vapour bubbles and result in maximum separation performance.
Contol Unit C-620 - A must for Automated Systems
Control Unit C-620:
The Control Unit C-620 in combination with SepacoreControl provides precise control of the chromatography system. The following components can be connected to the Control Unit C-620:
- 2 to 4 Pump Modules C-601 or C-605
- Up to 2 Fraction Collectors
- Up to 8 Detectors e. g. UV, RI
- Sequential Modules C-623 or C-625 for automatic sequential chromatography on up to 5 columns or cartridges
- The Control Unit C-620 is included in the SepacoreControl package
Automated System:
The fully automated chromatography system including UV-detection, fraction collector, column valves and software amongst many more features can be built up according to your specific separation needs.
Pump Modules for a wide range of applications
Pump Module C-601
The optimal selection for isocratic Flash separations: Silent operating 3-piston Pump Module C-601 for Flash Chromatography up to 10 bar. The pump module provides a constant, pulse-free flow from 2.5 to 250 ml/min and ensures reproducible, fast separation at a maximum working pressure of 10 bar/145 psi. For sample sizes of up to 5 g, pre-packed PP cartridges can be used for the quick, safe implementation of normal phase and reversed phase applications.
Fast separation with reversed phase and large sample size
Pump Module C-605
Similar to the Pump Module C-601 but with a maximum working pressure of 50 bar/725 psi. Using the Pump Module C-605, fast separation with reversed phase and separations can be performed with sample sizes up to 100 g. Ideal for use with glass and plunger columns and silica gel particle sizes < 40 μm.